Product Description:
The MT180 is a multi-mode ultrasonic thickness gauge, capable of measuring the thickness of various
material types with a high degree of accuracy. The multi-mode feature of the gauge allows the user to
toggle between pulse-echo mode for normal thickness measurement, and echo-echo mode which will
eliminate any paint or coating thickness from the reading.
Features:
· Multi-mode: Pulse-Echo mode and Echo-Echo mode.
· Capable of performing measurements on a wide range of material, including metals, plastic,
ceramics, composites, epoxies, glass etc.
· Various probes (transducers) are available for special application requirements.
· One and two-point calibration function.
· Single point mode and Scan mode. Seven measurements readings per second in single point
mode, and sixteen per second in Scan Mode.
· Coupling status indicator showing the coupling status.
· Units: metric or imperial readout is selectable.
· Battery capacity indicator.
· Auto sleep and auto power off function to conserve battery life.
· Optional software to process the memory data on the PC.
Configuration:
· The standard configuration includes the instrument, P5EE 5 MHz probe, 2 AA alkaline
batteries, couplant, manual and carry case.
· Options include the standard 2 MHz, 5 MHz and 7 MHz probes, the 5 MHZ HT high temp
probe, and the download software and cable.
Specifications:
· Range in pulse-echo mode: 0.65~600 mm in steel.
· Range in echo-echo mode: 3~30 mm in steel.
· Sound Velocity Range: 1000~9999 m/s.
· Resolution: 0.1 mm/0.01 mm.
· Accuracy: ± 0.5% of the thickness reading + 0.01 mm depending on material type and
conditions.
· Memory storage of up to 20 files, with up to 99 values per file.
· LCD display.
· Two “AA” size, 1.5 Volt alkaline batteries allowing for up to 100 hours typical operating time.
· Dimensions:150 mm×74 mm×32 mm.
· Weight:245 grams.
Operating Conditions:
· Operating Temperature: -20°C~+ 60°C;
· Storage Temperature: -30°C~+ 70°C
· Relative Humidity ≤90%;
· The surrounding environment should be free of vibration, strong magnetic fields, corrosive
materialsand heavy dust
Transducer Selection:
Model Freq
(MHz
Ещё видео!