Today's video is on how to calibrate and bump test the MSA Altair 5X unit. This uses a demand flow regulator and the standard MSA 4-gas mix. Remember to subscribe and like the video to help us out.
If you have a problem during your calibration, give us a call or send it in for repair here:
[ Ссылка ]
Calibration steps:
1. Turn monitor on
2. Complete pump test
3. Wait for startup to finish, hit NO when it asks to FAS
4. Press and hold right button
5. On prompt to zero, hit the left button for YES
6. When zero completes, hit YES when it asks you to span cal
7. Hook the gas up to the monitor
8. Wait for the calibration to complete
9. Remove the gas, you're done!
Calibration Gas Used:
20ppm H2S
60ppm CO
58% LEL Pentane Simulant
15.0% VOL O2
Nitrogen Balance