Chapter 16. Pause times in practices and clinics
The high-performance MELAtherm 10 washer-disinfector cleans and disinfects without interruption. The MELAtherm 10 tutorial shows you what to do during an operating pause.
Should the washer-disinfector not be used for two days or longer, start the Rinse-Program twice before instrument decontamination. In ophthalmology clinics with the corresponding instruments, we recommend the start of the Ophthalmo-Program.
Given a pause time of over 2 weeks, prepare the washer-disinfector as in the following fashion: Place the suction lances in a container with water and start the air-removal program. You can then replace the suction lances in the washer-disinfector. Please Check that the washing chamber is dry and switch off the power and water supplies.
Recommission the washer-disinfector after an operating pause by starting the air-removal program and the disinfection program (e.g. a Universal-Program) without instruments.
Learn more: [ Ссылка ]
#washer-disinfector #thermal-disinfector #MELAtherm